临床麻醉中混合静脉血氧饱和度监测具有重要的作用,需要通过置入的肺动脉导管进行测量,该方法属于有创方法,具有一定的并发症和风险,同时需要一定的临床操作经验,下面介绍一种无创的连续监测方法,当然了,还处于动物实验验证阶段,期待能用于临床的,无创的,简便的监测方法。



一种连续无创测量混合静脉血氧饱和度的方法:一项对猪的概念验证研究


背景:混合静脉血氧饱和度(SvO2)在评价供氧与全身耗氧平衡时非常重要。到目前为止,监测SvO2需要肺动脉导管采集血样。本文将容积式容量二氧化碳描记术(volumetric capnography,测量有效的肺血流的方法)与Fick氧耗量原理相结合,建立了一种连续无创的capnodynamic SvO2法来测量SvO2。这项研究的目的是通过SvO2测量金标准来验证这项新技术,通过肺动脉导管获取血样是心输出量(CO)-氧饱和度是SvO2测量的(CO-oximetry SvO2)金标准,并评估肺内分流的潜在影响。


方法:8头麻醉机械通气的家养仔猪(平均体重23.9kg)接受一系列旨在降低和增加SvO2的干预措施。在整个研究方案中,同时记录capnodynamic SvO2和CO-oximetry SvO2以及分流分数,使用Berggren公式。采用Bland-Altman图和协方差分析法评估capnodynamic SvO2和CO-oximetry SvO2绝对值的一致性以及capnodynamic SvO2检测变化的能力。


结果:capnodynamic SvO2与CO-oximetry SvO2的总偏差为-1个百分点(一致性限为-13至+11个百分点),平均百分误差为22%,一致性为100%。分流分数在基线时为13%,在研究结束时为22%,并且测试方法之间的一致性只有微小的变化。


结论:在目前的实验环境下,SvO2的capnodynamic法测量值的绝对值非常接近于CO-oximetry法的测量值,并且有100%的趋势分析能力。


A Continuous Noninvasive Method to Assess Mixed Venous Oxygen Saturation: A Proof-of-Concept Study in Pigs


Background: Mixed venous oxygen saturation (SvO2) is important when evaluating the balance between oxygen delivery and whole-body oxygen consumption. Monitoring SvO2 has so far required blood samples from a pulmonary artery catheter. By combining volumetric capnography, for measurement of effective pulmonary blood flow, with the Fick principle for oxygen consumption, we have developed a continuous noninvasive method, capnodynamic SvO2, for assessment of SvO2. The objective of this study was to validate this new technique against the gold standard cardiac output (CO)-oximetry SvO2 measurement of blood samples obtained from a pulmonary artery catheter and to assess the potential influence of intrapulmonary shunting.


Methods: Eight anesthetized mechanically ventilated domestic-breed piglets of both sexes (median weight 23.9 kg) were exposed to a series of interventions intended to reduce as well as increase SvO2. Simultaneous recordings of capnodynamic and CO-oximetry SvO2 as well as shunt fraction, using the Berggren formula, were performed throughout the protocol. Agreement of absolute values for capnodynamic and CO-oximetry SvO2 and the ability for capnodynamic SvO2 to detect change were assessed using Bland-Altman plot and concordance analysis.


Results: Overall bias for capnodynamic versus CO-oximetry SvO2 was -1 percentage point (limits of agreement -13 to +11 percentage points), a mean percentage error of 22%, and a concordance rate of 100%. Shunt fraction varied between 13% at baseline and 22% at the end of the study and was associated with only minor alterations in agreement between the tested methods.


Conclusions: In the current experimental setting, capnodynamic assessment of SvO2 generates absolute values very close to the reference method CO-oximetry and is associated with 100% trending ability.


 

免责声明:

本微信公众平台所刊载原创或转载内容不代表米勒之声的观点或立场。文中所涉及药物使用、疾病诊疗等内容仅供参考。

本文由作者自行上传,并且作者对本文图文涉及知识产权负全部责任。如有侵权请及时联系(邮箱:guikequan@hmkx.cn
关键词:
饱和度,血氧,静脉,测量,监测

人点赞

收藏

人收藏

打赏

打赏

我有话说

0条评论

0/500

评论字数超出限制

表情
评论

为你推荐

推荐课程


社群

精彩视频

您的申请提交成功

确定 取消
剩余5
×

打赏金额

认可我就打赏我~

1元 5元 10元 20元 50元 其它

打赏

打赏作者

认可我就打赏我~

×

扫描二维码

立即打赏给Ta吧!

温馨提示:仅支持微信支付!